The steps below provide a quick check of the calibration of the MultiClamp 700B.  It is assumed that appropriate shielding (as described in “Test the Noise", above) will be used during these tests.

1.  Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.

2.  Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults" button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.


50 G Range

1.  Press the “Options" button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane.  Return to the main MultiClamp 700B Commander window.

2.  Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.

3.  Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.

4.  Press Auto Cp Fast to remove the bulk of electrode capacitance transient.

5.  The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms.  The rise time to the peak of the overshoot should be about 50 μs.  The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).


5 G Range

1.  Change the feedback resistor to 5 G in the Options / Gains menu.

2.  Press Auto Cp Fast.

3.  The step response should be ~50 mVp-p (±5 mV).


500 M Range

1.  Press the “Reset to Program Defaults" button.  (By default, the 500 M

range is selected.)

2.  Check Seal Test and set the amplitude to 25 mV.

3.  Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.

4.  Press the Auto Whole Cell button.

5.  Press Auto Cp Fast button.

6.  The step response should be ~25 mVp-p. 


50 M Range

1.  Change the feedback resistor to 50 M.

2.  Increase Output Gain to 10.

3.  Press Auto Whole Cell and Auto Cp Fast.

4.  The step response should be ~25 mVp-p.  



[ 打印 ] [ 返回顶部 ] [ 关闭

| 商铺首页 | 公司档案 | 产品展示 | 供应信息 | 公司动态 | 询价留言 | 联系我们 | 会员管理 |
化工仪器网 设计制作,未经允许翻录必究.Copyright(C) http://www.mtfinger.com, All rights reserved.
以上信息由企业自行提供,信息内容的真实性、准确性和合法性由相关企业负责,化工仪器网对此不承担任何保证责任。
温馨提示:为规避购买风险,建议您在购买产品前务必确认供应商资质及产品质量。
二维码

扫一扫访问手机站
包头市| 肃南| 陕西省| 恩施市| 汝阳县| 丘北县| 涞源县| 铜陵市| 吴桥县| 阳朔县| 三原县| 仲巴县| 扎赉特旗| 福州市| 永州市| 陆川县| 大田县| 门源| 日土县| 闵行区| 湖州市| 乃东县| 临西县| 乌海市| 云安县| 渝中区| 柳林县| 台南县| 合山市| 绥宁县| 乌恰县| 阜城县| 宜都市| 贡山| 广河县| 红原县| 衡南县| 阜南县| 湘西| 蓬溪县| 延川县|